Automatic Inspection (Sequence Program Function)
3
If [ON] is selected, set the save number.
0 to 9 (default setting: 0)
Press the F3 key [↓] or F4 key [↑] to select a number.
4
Set the discharge time before measurements start.
Press F3 key [ms/s] to switch the unit.
0 ms s to 999.9 s (default setting: 0.0 s)
Pressing the F4 key
[EDIT] enables the
numerical value to be
changed.
If the setting of the voltage output function for no voltage output is set to [DISC (D)], the
measurement object is discharged while no measurement is performed. The instrument will not
discharge voltage if it is disconnected from the measured object by external relays or if the setting of
the voltage output function for no voltage output is set to high impedance ([Hi-Z]).
In such case, set Discharge 1 ([DCHG1]). Also, set the setting to Discharge 1 ([DCHG1]) when
electric charge is remaining in the measured object due to repeated measurement.