Automatic Inspection (Sequence Program Function)
4.3 Automatic Inspection (Sequence Program
Function)
Be sure to read “Before removing the measured object” (p. 13) beforehand.
After a test pattern including discharge, charge, measurement and discharge is set, the tests can
be executed in sequence (sequence measurement).
Thus, the time of applying the measurement voltage to the object to be measured can be controlled
accurately.
The sequence measurement can inspect objects under the equal condition by setting the following
items rst.
[DCHG1] Discharge 1 Sets the discharge time before measurements start.
[CHARG] Charge Sets the charging (voltage application) time before
measurements start.
[MEAS] Measurement Sets the measurement time. The measured value is output at
this time.
[DCHG2] Discharge 2 Sets the discharge time after measurements complete.
Set them on the Measurement screen.
1
Press the F4 key [MODIFY] to enable the measurement conditions to be changed.
If no items are displayed, press F3 key [MONI] to switch the measurement condition display.
Pressing the ESC key cancels the changes of the measurement conditions.
2
Select whether to perform sequence measurement or not.
[ON] Yes
[OFF] No (default setting) (Measurement will be done continuously.)