Settings Common to Interfaces
10.5 Settings Common to Interfaces
Outputting measured values (data output function)
(RS-232C and USB only)
Select whether to output the measured values automatically via each of the interfaces.
Setting this function to [ON] disables the communication commands to control the instrument
because the data output exclusively uses the interface.
Procedure to display the setting screen: MENU key >
[IF] tab
[ON] Outputs measured values.
[OFF] Does not output any measured values. (default setting)
For the external trigger setting:
Measured values are output when the TRIG signal is input to the instrument or when the
ENTER (TRIG) key is pressed.
For the internal trigger setting:
Latest measured values are automatically output every time the measurement is completed.