43
Setting Measurement Speed
3.3 Setting Measurement Speed
The slower measurements are, the more the accuracy improves.
Pressing the SPEED key on the measurement screen changes the measurement speed.
See “Current measurement accuracy” (p. 141).
Some measurement speeds are not available depending on the current range setting. If the
measurement speed cannot be changed, check the current range.
See “Available setting depending on current ranges and measurement speed” (p. 43).
Press the SPEED key on the measurement screen.
Pressing the F4 key [MODIFY] enables also the function keys to be used to select the
measurement speed.
FAST, FAST2, MED, SLOW, SLOW2 (Default setting: SLOW2)
The measurement speed changes every time the F3 key [↓] is pressed in the following order:
FAST, FAST2, MED, SLOW, and SLOW2.
The measurement speed changes every time the F4 key [↑] is pressed in the following order:
SLOW2, SLOW, MED, FAST2, and FAST.
Available setting depending on current ranges and measurement speed
: Setting available, –: Setting not available
Measurement
speed
Range
FAST FAST2 MED SLOW SLOW2
20 pA – – –
200 pA – –
2 nA
20 nA
200 nA
2 µA
20 µA
200 µA
2 mA
– – –
3
Basic Measurement