Function specications
Measurement speed Operation details Sets the sampling time.
Setting
FAST/FAST2/MED/SLOW/SLOW2
Range switching
function
Current measurement
AUTO/MANUAL
Resistance
measurement
No setting (The current measurement set to AUTO will set the
resistance measurement to AUTO.)
Trigger mode Operation details Sets a trigger for starting measurement.
Setting
INTERNAL/EXTERNAL
INTERNAL Internal trigger
EXTERNAL External trigger Pressing the TRIG key samples a
datum once.
Delay function Operation details Sets the duration from when the START key is pressed or the
TRIG signal is input until the start of measurement.
Setting range
0 ms to 9999 ms (in increments of 1 ms)
Contact check delay
function
Operation details Sets the duration before starting contact check
Setting range
0 ms to 9999 ms (in increments of 1 ms)
Average function Operation details Averages measured values.
Setting
OFF/ON (HOLD)/ON (AUTO)
Number of averaging
with averaging set to
ON
2 to 255 times (Any value in this range can be set.)
Averaging method
HOLD Moving average (However, arithmetic averages are
calculated with the trigger source set to the external
trigger and the sequence measurement set to o󰀨)
AUTO The number of times of averaging is automatically
changed based on the variation of the measured values.
Power frequency
setting
Operation details Sets the power voltage frequency.
Setting
AUTO (Automatically detects whether the frequency is 50 Hz or
60 Hz)/50 Hz/60 Hz
Cable length
correction function
Operation details The predetermined length of the cable to be used corrects the
xture capacitance open value and the contact check value.
Automatically detected length and user-defined length can
be used.
Correctable range 0.5 m to 3.0 m, default: 1.0 m (Can be set in increments of 0.1 m)
Fixture capacitance
open correction
function
Operation details Measures the capacity when the xture is open.
Required to be executed prior to executing the contact check
function.
The acquired data is retained even when the instrument is turned
o󰀨.
Display range 0.000 pF to 99.999 pF