EasyManua.ls Logo

Hioki Super megohm SM7110 - Using Instrument as Ammeter (If Operated in Combination With External Power Supply)

Hioki Super megohm SM7110
194 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Appx.7
Using Instrument as Ammeter (If Operated in Combination With External Power Supply)
Appx. 5 Using Instrument as Ammeter (If Operated
in Combination With External Power
Supply)
Be sure to read “Before using the instrument as an ammeter” (p. 14) beforehand.
1
Connect the object to be measured and the other devices to the instrument as follows.
A
INPUT
GUARD
GROUND
OUTPUT
R R
p
V
o
SM7110/SM7120
+
Protective
resistor
External power supply
Object to be
measured
Short-circuit
Object to be
measured
Protective
resistor
External power supply
Short-circuit GROUND and
GUARD with shorting plug
Measurement
voltage source
Current limiter
Ammeter
When using the external power supply, insert a protective resistor so that a current owing through
the instrument is reduced to the current rating or less even if the terminals of the object to be
measured are short-circuited with each other.
Test using a measurement voltage of 1000 V or less: 50 mA or less
Test using a measurement voltage exceeding 1000 V: 1.8 mA or less
2
Congure the instrument setting as follows:
Trigger: Internal trigger (INT) (p. 58)
Setting of voltage value for resistance calculation: voltage set value (set value is external
power supply output value) (p. 56)
3
Turn on the external power supply.
4
Connect the measurement leads to the object to be measured.
5
Apply the voltage from the external power supply.
Appx.

Table of Contents