Appx.13
Modifying Measurement Lead
Appx. 7 Modifying Measurement Lead
When modifying the tip of the Hioki-made measurement lead, follow the procedure below:
When stripping o󰀨 the jacket or braided conductor, take care not to break the wires or short-circuit them with
each other.
We will not guarantee the accuracy of measured values obtained using measurement leads modied by users.
Structure of shielding wire of Hioki-made measurement leads
Jacket
Braided conductor for the shielding wire
Insulator
Braided conductor for the guard wire
Conductive vinyl
Conductor
Modifying the tip of measurement leads
1
Strip o󰀨 the jacket from with a knife, etc.
Braided conductor for the shielding wire
Jacket
2
Turn the braided conductor for the shielding wire inside out.
InsulatorBraided conductor for the shielding wire
3
Strip o󰀨 the insulator with a knife, etc.
Braided conductor for the guard wireInsulator
4
Turn the braided conductor for the guard wire inside out.
Keep them away from each other not to make a short-circuit.
Take care not to short-circuit the shielding wire and the guard wire with each other.