24
Displayed Contents of Measurement Screen
1.4 Displayed Contents of Measurement Screen
1 2 3 4 5 7
8
9
10
11
12
14
13
6
Displays menu settings screen (p. 27)
Performs contact check.
(Available only with contact
check setting set to [ON])
Changes measurement condition
contents to be displayed.
Enables change of the
measurement conditions
Name Description
1
Trigger setting
Displays the presently set trigger.
[INT] Internal trigger
[EXT] External trigger
2
Sampling data Appears during the sampling data.
3
Self-calibration Appears during the self-calibration.
4
Voltage monitor check
Appears when the voltage monitor check function is enabled.
Highlighted in reverse video if the voltage is abnormal.
5
Contact check
Appears when the contact check function is enabled.
Highlighted in reverse video if a contact error occurs.
6
Open correction
execution result
Displays the result when the open correction is executed once.
7
Interlock
Key lock
[I.LOCK] Appears while the interlock function is activated.
[K.LOCK] Appears while the key lock function is activated.
Displays [I.LOCK] while both the interlock and the key lock functions are
activated.
Remote [RMT]
Appears while the instrument is placed in the remote status
(communicating with an external device).
Pressing the LOCAL key disables the remote control.
8
Measurement judgment Displays the judgment result if the comparator function is enabled.
9
Voltage monitor value The voltage produced by the instrument is displayed.
10
Temperature
Displays temperature and humidity if the Humidity Sensor is connected.
Displays an error message if the Humidity Sensor is not connected to the
instrument (p. 161).
See “2.4 Connecting Humidity Sensor” (p. 31).
11
Humidity
12
Measured value
Displays the measured value corresponding to the measured value display
mode (p. 39).