68
Contact Check (Various Setting)
Setting of [WORK.C]
Set the capacitance of the object to be measured.
[LOW] If the capacitance of the objects to be measured is 10 pF or less
[NORMAL] If the capacitance of the objects to be measured exceeds 10 pF (default setting)
If the capacitance of the object to be measured is unknown, select [NORMAL] and then execute the
open correction with the object connected to the instrument to obtain its capacitance.
See “4.11 Canceling Capacitance of Measurement Fixture (Fixture Capacitance Open Correction
Function)” (p. 70).
Setting of [CABLE]
Set the cable length.
[AUTO] The cable length is automatically detected and set.
[EDIT] Optional numerical values can be set.
0.5 m to 3.0 m (Can be set in increments of 0.1 m, default setting: 1.0 m)
Pressing the F4 key
[EDIT] enables the
numerical value to be
changed.
If using an accessory or an optional cable, press the F4 key [EDIT] to specify its length.
If the cable is modied or extended, it is recommended to press the F3 key [AUTO].
Pressing the F3 key [AUTO] executes the cable length correction once. If you change the
cables, press the F3 key [AUTO] to execute the cable length correction again.