72
Contact Check (Executing Contact Check, Setting Reference Value)
Procedure to display the settings screen: (Measurement screen) MENU key > [C.CHK] tab
1
Select whether to perform contact check or not.
[ON] Enabled (The contact check is executed before starting measurement.)
[OFF] Disabled (default setting)
Executes contact check.
2
Set the capacitance reference value for judgment.
Set the minimum capacitance value for measurement objects.
Judgments are delivered by comparing measured values to this reference value.
A contact error occurs if the capacitance is less than this value.
Pressing the F4 key
[EDIT] enables the
numerical value to be
changed.
Settle
Cancel
Move to the next
digit (left or right)
Change the
numerical value
(up or down)
Numerical value can also be changed with the numeric keypad. See “Methods for changing
numerical values” (p. 28).