Service Manual
Storm³
As at: 07.03.02 Page 15
replace drive unit /
check carbon
brushes
Drive unit
defective
Wheelchair judders in drive mode
chapt. 8.3.2
replace battery
replace charging
device
Batteries defective
Charging device defective
Cannot charge batteries
chapt. 9.3.1
Warning horn defective
replace remote
Lighting defective
check / replace
cabling or bulbs
chapt.
11.1.4
chapt. 10
Electrical leg support defective
check / replace
cabling or motor
Manual backrest adjustment defective
check / replace
gearbox, handwheel,
handwheel shaft,
armature shaft
chapt. 14.5
chapt.
13.1.21
Electrical backrest adjustment defective
check / replace
cabling, remote,
motor / check limit
switch
chapt. 13.3
chapt.
13.3.2
Wheelchair runs too slowly
replace contact pins
Contact pins in seat frame defective (legrest
attachment)
Chapt.
11.4.2
Remote defective
Power module defective
Lifter not at bottom (2x blink)
replace remote
replace power
module
lower lifter / reed
sensor defective /
magnet missing
chapt.
11.1.4
chapt.
11.4.2
chapt. 9.2.2
(note)