EasyManua.ls Logo

Invacare Storm 3 - 8.3.2 Replacing drive unit, pressure spring; 8.3.3 Replacing carbon brushes

Invacare Storm 3
93 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Service Manual
Storm³
As at: 07.03.02 Page 25
- 1 Z131035 SPARE PART BRAKE KIT RH FOR DRIVE P01061/3
(SCHMID)
- 1 Z131034 SPARE PART BRAKE KIT LH FOR DRIVE P01062/4
(SCHMID)
- 1 P09292 BRAKE COVER FOR DRIVE (02 SCHMID)
- 1 SP1423439 CARBON BRUSH SET SCHMID
- 2 P09293 CARBON BRUSHES FOR DRIVE (02 SCHMID)
- 1 F351007 ACS MOTOR PLUG ASSY.
8.3.2 Replacing drive unit, pressure spring
Remove taxi strap if necessary by unscrewing all four
screws.
Remove battery cover (see page 39, chapt. 9.2.1).
Remove the respective motor plug.
Unscrew all plastic adjusting screws and dismantle the
complete equipment rack (see page 54, chapt. 11.4.1).
Dismantle disengage lever (see page 27, chapt. 0?).
Remove batteries (see page 39, chapt. 9.2.1).
Dismantle the respective rear wheel (see page 36, chapt.
9.1.7).
Dismantle the respective bulb holder (see page 43, chapt.
10.2.2).
Pass a suitable screw through the diaphragm (1) to
pretension the spring (M8 x 120).
Unscrew the hex nut (2) and drive out the Allen key.
Remove the drive unit with the rocker block (3) downwards
and replace if necessary.
Slacken the defective pressure spring (4) if necessary and
replace.
Reassembly takes place in the reverse order.
8.3.3 Replacing carbon brushes
Note:
When replacing carbon brushes, ensure that the spring located behind the plastic cap
does not get lost.
"

Table of Contents

Other manuals for Invacare Storm 3

Related product manuals