Storm³
Service Manual
Page 84 As at:07.03.02
A = Brown, B = Blue
• Unscrew cylinder head screws (6. 7) and hex bolt (8).
• Cut off cable binders.
• Pull nut (10) out of the tube profile.
• Pull protector (15. 16) out of the tube profile.
• Lift motor (1) can be replaced.
• Ensure that the contacts are correctly located with
reassembling.
• Plug the cable into the corresponding recesses (see
diagram left).
• Push protector (15. 16) into the tube profile.
• Fit leg support.
• Fix cable with cable binder.
• Check leg support for correct functioning.
• If the travel direction is incorrect, the legrest contact (1) in
the footrest adapter can be turned through 180°.
Reassembly takes place in the reverse order.