Page 84 As at:07.03.02
A = Brown, B = Blue
Unscrew cylinder head screws (6. 7) and hex bolt (8).
Cut off cable binders.
Pull nut (10) out of the tube profile.
Pull protector (15. 16) out of the tube profile.
Lift motor (1) can be replaced.
Ensure that the contacts are correctly located with
reassembling.
Plug the cable into the corresponding recesses (see
diagram left).
Push protector (15. 16) into the tube profile.
Fit leg support.
Fix cable with cable binder.
Check leg support for correct functioning.
If the travel direction is incorrect, the legrest contact (1) in
the footrest adapter can be turned through 180°.
Reassembly takes place in the reverse order.