Keysight B2961A/B2962A User’s Guide, Edition 3 4-7
Front Panel Reference
Display and Assist Keys
Dual View
Available on B2962A. The upper half is for channel 1. The lower half is for channel
2. Each area displays the measurement results, source and measurement settings,
and channel status. Most setup parameters can be edited on this display.
Display area 1. Channel number. 1 or 2.
2. Output status (OFF, CV, CC, or EMUL)
CV: Constant Voltage
CC: Constant Current
EMUL: Emulation Table (this is displayed when programmable output
resistance is set using the emulation table.)
3. Over voltage/current status indicator (OV or OC, red)
4. Latest measurement data (4½ digit resolution)
5. High capacitance mode indicator (HC)
6. Programmable output resistance indicator
7. External filter indicator
8. Remote sensing (4-wire connection) status indicator
9. Source function. VOLTS or AMPS.
10. Source shape indicator. DC, waveform, pulse, sweep, or pulsed sweep. DC does
not show the indicator.