Keysight B2961A/B2962A User’s Guide, Edition 3 5-19
Front Panel Operations
To Set Source Output
To cancel the setting change, press the Cancel/Local key.
To Enable or Disable Over Voltage/Current Protection
For details on this function, see “Over Voltage/Current Protection” on page 6-14.
1. Press the Config > Source > Connection function keys.
The Output Connection dialog box opens.
2. In the Ch field, set the channel to which the setting change is applied.
3. Set ON (enable) or OFF (disable) in the Over Voltage/Current Protection field.
4. To apply the settings, press Apply.
To apply the settings and close the dialog box, press OK.
To cancel the setting change, press the Cancel/Local key.
To Specify Output-Off Status
For details on this function, see “Output-Off Status” on page 6-15.
1. Press the Config > Source > Connection function keys.
The Output Connection dialog box opens.
2. In the Ch field, set the channel to which the setting change is applied.
3. Set HIGH Z (high impedance), NORMAL (normal), or ZERO (zero volt) in the
Output-Off State field.
4. To apply the settings, press Apply.
To apply the settings and close the dialog box, press OK.
To cancel the setting change, press the Cancel/Local key.
To Enable or Disable Automatic Output-On Function
For details on this function, see “Automatic Output-ON/OFF Function” on page
6-16.
1. Press the Config > Source > Connection function keys.
The Output Connection dialog box opens.
2. In the Ch field, set the channel to which the setting change is applied.