Keysight B2961A/B2962A User’s Guide, Edition 3 4-29
Front Panel Reference
Config key group
Set the function ON to turn on the channel
output automatically just before the trigger
system is initiated by the SCPI command (not
by front panel operation).
Auto Output-Off Automatic output off function, ON or OFF
Set the function ON to turn off the channel
output automatically and immediately when all
trigger system changes the status from busy to
idle.
Output Filter dialog box
This dialog box provides the following parameters for setting the output filter.
Ch Only on 2-channel models. Channel 1 (Ch 1) or 2 (Ch 2)
This field specifies the channel set by this dialog box.
Filter State Output filter, ON or OFF
Set the filter ON to obtain clean source output without spikes
and overshooting. Note, however, that using a filter may
increase the channel settling time.
Automatic Filter Automatic filter, ON or OFF
Set the function ON to automatically set the output filter which
provides the best filter characteristics and cutoff frequency.
Time Constant Filter time constant, 5 s to 50 ms
External Filter dialog box
This dialog box provides the following parameters for using the
N1294A-020/021/022 low noise filter. The filter cannot be detected automatically.
The following settings are saved to the non-volatile memory.
Ch Only on 2-channel models. Channel 1 (Ch 1) or 2 (Ch 2)
This field specifies the channel set by this dialog box.
External Filter
State Use of the filter, ON for use or OFF for non-use