Keysight B2961A/B2962A User’s Guide, Edition 3 6-3
Function Details
Limit/Compliance
Limit/Compliance
Limit/compliance is the output limiter for preventing damage to the test device from
overcurrent or overvoltage. Voltage limit is for current output channels, and current
limit is for voltage output channels.
When a channel reaches limit, the channel acts as a constant voltage source or a
constant current source. The channel keeps the output value once limit has been
reached.
Limit can be set with the same resolution and accuracy as output current or output
voltage.
To Set Limit
The following points must be noted when setting the limit.
The minimum limit value is limited as follows.
Current limit
1 % of the range (at 100 nA Range)
1 nA (at 10 nA Range)
Voltage limit
20 mV (at 0.2 V range)
If the current limit value is too low, the channel will require a long settling time.
The limit value can be specified asynchronously to positive and negative.

Table of Contents

Other manuals for Keysight Technologies B2961A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies B2961A and is the answer not in the manual?

Keysight Technologies B2961A Specifications

General IconGeneral
BrandKeysight Technologies
ModelB2961A
CategoryPortable Generator
LanguageEnglish

Related product manuals