5-28 Keysight B2961A/B2962A User’s Guide, Edition 3
Front Panel Operations
To Use Trace Buffer
To Display Statistical Data
For details on the setup parameters, see “Trace Statistical Result dialog box” on
page 4-39.
1. Press the Result > Trace function keys.
The Trace Statistical Result dialog box opens.
2. In the Ch field, set the channel for the data to be displayed.
3. In the Element field, set the data type SOURCE (source output data), VOLTS
(voltage measurement data), AMPS (current measurement data), OHMS
(resistance measurement data) for the statistical calculation.
The statistical data for the data specified here is displayed in the area below the
Length field.
4. Press OK to close the dialog box.
The Length field displays the number of data.
The statistical data is displayed in the following fields:
Mean: Mean value
Std. Dev.: Standard deviation
Min.: Minimum value
Max.: Maximum value