Loading...
6-16 Keysight B2961A/B2962A User’s Guide, Edition 3
Function Details
Automatic Output-ON/OFF Function
Automatic Output-ON/OFF Function
The automatic output-ON/OFF function defines the source channel’s output
ON/OFF operation when the trigger system changes the status.
Automatic-ON function
If this function is enabled, the channel automatically turns the output ON just
before the trigger system is initiated by an SCPI command, not by a front panel
operation.
Automatic-OFF function
If this function is enabled, the channel automatically turns the output OFF
immediately when all trigger system changes the status from busy to idle.
To set the automatic output ON/OFF function, see “Output Connection dialog box”
on page 4-28.