3-6 Keysight B2961A/B2962A User’s Guide, Edition 3
Installation
Installing the Keysight B2961A/B2962A
Installing the Keysight B2961A/B2962A
This section describes information you must pay attention to when installing the
Keysight B2961A/B2962A.
• “Safety Considerations”
• “Environment”
• “Connecting the Power Cord”
• “Setting the Power Line Frequency”
• “Setting Date and Time”
• “Bench Installation”
• “Rack Installation”
Safety Considerations
Refer to the Safety Summary page at the beginning of this guide for general safety
information. Before installation or operation, check the B2961A/B2962A and
review this guide for safety warnings and instructions. Safety warnings for specific
procedures are located at appropriate places throughout this guide.
Environment
WARNING Do not operate the instrument in dusty environment, or in the presence of
flammable gases, corrosive gases, or fumes.
Ne pas utiliser l’appareil dans un endroit poussiéreux, ou en présence de gaz
inflammables, corrosifs ou de fumée.
Environmental conditions for the B2961A/B2962A are described below. The
B2961A/B2962A is designed for use in indoor facilities.
• Temperature range
Operating 0 C to +55 C
Storage -30 C to +70 C
• Humidity range
Operating 30 % to 80 % RH, non-condensing