4-34 Keysight B2961A/B2962A User’s Guide, Edition 3
Front Panel Reference
Function key group
Trace Buffer Setup dialog box
This dialog box provides the following parameters for setting the trace function. The
data specified by the Feed Data parameter can be stored in the trace buffer if the
Buffer Control parameter is NEXT. The maximum data size is specified by the
Buffer Size parameter. See Figure 6-6 on page 6-23 for the trace buffer.
Ch Only on 2-channel models. Channel 1 (Ch 1) or 2 (Ch 2)
This field specifies the channel set by this dialog box.
Feed Data Type of data placed in the trace buffer, SENSE or MATH
SENSE: Measurement result data
MATH: Calculation result data of math expression
The data contains the voltage measurement data, current
measurement data, resistance measurement data, source output
setting data, calculation result data, time data, or status data
selected by using the Format keys of the I/O key group. See
“I/O key group” on page 4-42.
Buffer Control Trace buffer control mode, NEVER or NEXT
NEVER: Disables the write operation to the trace buffer.
NEXT: Enables the write operation until buffer full.
Buffer full will change the mode to NEVER.
Buffer Size Size of the trace buffer, 1 to 100000 data

Table of Contents

Other manuals for Keysight Technologies B2961A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies B2961A and is the answer not in the manual?

Keysight Technologies B2961A Specifications

General IconGeneral
BrandKeysight Technologies
ModelB2961A
CategoryPortable Generator
LanguageEnglish

Related product manuals