4-32 Keysight B2961A/B2962A User’s Guide, Edition 3
Front Panel Reference
Config key group
Ch Only on 2-channel models. Channel 1 (Ch 1) or 2 (Ch 2)
This field specifies the channel set by this dialog box.
State Wait time, ON or OFF
Automatic Automatic wait time, ON or OFF
Gain and Offset Parameters for calculating the wait time. See the following
formula.
• If State=ON and Automatic=ON:
wait time = Gain initial wait time + Offset
• If State=ON and Automatic=OFF:
wait time = Offset
• If State=OFF:
wait time=0