EasyManuals Logo

Keysight Technologies B2961A User Manual

Keysight Technologies B2961A
230 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #134 background imageLoading...
Page #134 background image
4-32 Keysight B2961A/B2962A User’s Guide, Edition 3
Front Panel Reference
Config key group
Ch Only on 2-channel models. Channel 1 (Ch 1) or 2 (Ch 2)
This field specifies the channel set by this dialog box.
State Wait time, ON or OFF
Automatic Automatic wait time, ON or OFF
Gain and Offset Parameters for calculating the wait time. See the following
formula.
If State=ON and Automatic=ON:
wait time = Gain initial wait time + Offset
If State=ON and Automatic=OFF:
wait time = Offset
If State=OFF:
wait time=0
The initial wait time is automatically set by the instrument, and cannot be changed.