EasyManuals Logo

Keysight Technologies B2961A User Manual

Keysight Technologies B2961A
230 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #139 background imageLoading...
Page #139 background image
Keysight B2961A/B2962A User’s Guide, Edition 3 4-37
Front Panel Reference
Trigger key group
Trigger Output Trigger output, ON or OFF
Trigger output terminal, an internal bus (INT1, INT2), LAN
port (LAN), or a digital I/O pin EXTn (n=1 to 14)
If this parameter is ON, B2961A/B2962A sends an output
trigger when it changes the trigger state for the action specified
by the Ch, Layer, and Action parameters. See Figure 6-7 on
page 6-27.
For using a digital I/O pin, it is necessary to set function of the
pin, output trigger polarity, type, pulse width, and so on. See
“DIO Configuration dialog box” on page 4-45 and “Using
Digital I/O” on page 3-37.
Table 4-6 Trigger Source
Trigger Source Description
AUTO Trigger source best suited for the present operating mode is
automatically selected by the internal algorithms.
BUS Remote interface trigger command such as GET, TRIGger,
and *TRG
TIMER Signal internally generated every interval set by the Period
field
INT1 or INT2 Signal from the internal bus 1 or 2
LAN LXI trigger specified by the
:ARM[:ACQ|:TRAN]:SOUR:LAN and
:TRIG[:ACQ|:TRAN]:SOUR:LAN commands
EXTn Signal from the DIO pin n, which is an I/O port of the Digital
I/O D-sub connector on the rear panel. n=1 to 14

Table of Contents

Other manuals for Keysight Technologies B2961A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies B2961A and is the answer not in the manual?

Keysight Technologies B2961A Specifications

General IconGeneral
BrandKeysight Technologies
ModelB2961A
CategoryPortable Generator
LanguageEnglish

Related product manuals