Keysight B2961A/B2962A User’s Guide, Edition 3 3-17
Installation
Connecting a DUT
Figure 3-4 Dual View, Floating Status for Channels 1 and 2
CAUTION Do not apply current to the chassis ground. Doing so will damage the instrument.
WARNING To prevent electrical shock, do not touch any of measurement circuit at any
time while a floating measurement is in progress. Also use accessories that
comply with IEC 61010-2-031. All terminals and the extended conductors must
be isolated by using insulation caps, sleeves, etc.
Afin d’éviter toute décharge électrique, ne touchez aucune mesure de circuit à
tout moment lorsque la mesure de flotte est en cours. Utiliser également des
accessoires qui sont conformes à la norme IEC 61010-2-031. Toutes les bornes
et les conducteurs prolongés doivent être isolés en utilisant des bouchons
d'isolation, des manchons, etc.