162 Keysight InfiniiVision 2000 X-Series Oscilloscopes User's Guide
10 Triggers
To trigger on all fields of the video signal
To quickly and easily view transitions between fields, or to find the amplitude
differences between the fields, use the All Fields trigger mode.
1 Press the [Trigger] key.
2 In the Trigger Menu, press the Trigger softkey; then, turn the Entry knob to
select Video.
3 Press the Settings softkey, then press the Standard softkey to select the
appropriate TV standard.
4 Press the Mode softkey and select All Fields.
Figure 29 Triggering on Field 1