Keysight InfiniiVision 2000 X-Series Oscilloscopes
User's Guide
15 Mask Testing
To create a mask from a "golden" waveform (Automask) / 227
Mask Test Setup Options / 229
Mask Statistics / 232
To manually modify a mask file / 233
Building a Mask File / 236
One way to verify a waveform's compliance to a particular set of parameters is to
use mask testing. A mask defines a region of the oscilloscope's display in which
the waveform must remain in order to comply with chosen parameters.
Compliance to the mask is verified point-by-point across the display. Mask test
operates on displayed analog channels; it does not operate on channels that are
not displayed.
Mask test is a license-enabled feature.
To create a mask from a "golden" waveform (Automask)
A golden waveform meets all chosen parameters, and it is the waveform to which
all others will be compared.
1 Configure the oscilloscope to display the golden waveform.
2 Press the [Analyze] key.
3 Press Features; then, select Mask Test.
4 Press Features again to enable mask testing.