248 Keysight InfiniiVision 2000 X-Series Oscilloscopes User's Guide
17 Waveform Generator
The Settings softkey opens the Waveform Generator Settings Menu which lets you
make other settings related to the waveform generator.
See:
"To output the waveform generator sync pulse" on page 248
"To specify the expected output load" on page 248
"To use waveform generator logic presets" on page 249
"To restore waveform generator defaults" on page 255
To output the waveform generator sync pulse
1 If the Waveform Generator Menu is not currently displayed on the
oscilloscope's softkeys, press the [Wave Gen] key.
2 In the Waveform Generator Menu, press the Settings softkey.
3 In the Waveform Generator Settings Menu, press the Trig Out softkey and turn
the Entry knob to select Waveform Generator Sync Pulse.
To specify the expected output load
1 If the Waveform Generator Menu is not currently displayed on the
oscilloscope's softkeys, press the [Wave Gen] key.
2 In the Waveform Generator Menu, press the Settings softkey.
Waveform Type Sync Signal Characteristics
All waveforms except
DC and Noise
The Sync signal is a TTL positive pulse that occurs when the waveform
rises above zero volts (or the DC offset value).
DC N/A
Noise N/A