Acquisition Control 12
Keysight InfiniiVision 2000 X-Series Oscilloscopes User's Guide 191
Infinite Persistence with Segmented Memory
When data has been acquired to segmented memory, you can also turn on infinite
persistence (in the Display Menu) and press the Analyze Segments softkey to create
an infinite persistence display. The Analyze Segments softkey appears when the
acquisition is stopped and the segmented memory feature is on.
Segmented Memory Re-Arm Time
After each segment fills, the oscilloscope re-arms and is ready to trigger in about
8µs.
Remember though, for example: if the horizontal time per division control is set to
5 µs/div, and the Time Reference is set to Center, it will take at least 50 µs to fill all
ten divisions and re-arm. (That is 25 µs to capture pre-trigger data and 25 µs to
capture post-trigger data.)
Saving Data from Segmented Memory
You can save either the currently displayed segment (Save Segment - Current), or all
segments (Save Segment - All) in the following data formats: CSV, ASCII XY, and
BIN.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies InfiniiVision 2000 X Series and is the answer not in the manual?

Keysight Technologies InfiniiVision 2000 X Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelInfiniiVision 2000 X Series
CategoryTest Equipment
LanguageEnglish

Related product manuals