Utility Settings 20
Keysight InfiniiVision 2000 X-Series Oscilloscopes User's Guide 285
• Mask— The pass/fail status is evaluated periodically. When the evaluation of
the testing period results in a failure, the trigger output pulses high (+5 V).
Otherwise, the trigger output remains at low (0 V). See Chapter 15, “Mask
Testing,” starting on page 227.
• Waveform Generator Sync Pulse— All of the waveform generator output
functions (except DC and Noise) have an associated Sync signal:
The Sync signal is a TTL positive pulse that occurs when the waveform rises
above zero volts (or the DC offset value).
See Chapter 17, “Waveform Generator,” starting on page 245.
The TRIG OUT connector also provides the User Cal signal. See "To perform user
calibration" on page 287.
Enabling Remote Command Logging
When remote command logging is enabled, remote commands sent to the
instrument (and results returned by the instrument) can be logged to the screen,
to a text file on a USB storage device, or to both the screen and a text file.
To enable remote command logging:
1 Press [Utility] > Options > Remote Log to open the Remote Log menu:
2 Press Enable to enable or disable the remote command logging feature.
When remote logging is enabled, additional debug information can be included
in the returned error string. If the error is detected by the SCPI command
parser, such as a header error or other syntax error, the extra debug
information is generated and included. But if the error is detected by the
oscilloscope system, such as when an out-of-range value is sent, then no extra
debug information is included.
3 Press Destination to select whether remote commands are logged to a text file
(on a connected USB storage device), logged to the screen, or both.
4 Press Write Mode to specify whether logged commands will be created in a new
list or appended to existing logged commands.