Vertical Controls 3
Keysight InfiniiVision 2000 X-Series Oscilloscopes User's Guide 67
To specify the probe skew
When measuring time intervals in the nanoseconds (ns) range, small differences in
cable length can affect the measurement. Use Skew to remove cable-delay errors
between any two channels.
1 Probe the same point with both probes.
2 Press one of the probes associated channel key.
3 In the Channel Menu, press Probe.
4 In the Channel Probe Menu, press Skew; then, select the desired skew value.
Each analog channel can be adjusted ±100 ns in 10 ps increments for a total of
200 ns difference.
The skew setting is not affected by pressing [Default Setup] or [Auto Scale].