CAN/LIN Triggering and Serial Decode 23
Keysight InfiniiVision 2000 X-Series Oscilloscopes User's Guide 341
If the total frame count exceeds the length specified in the standard, TFM will
appear red (LIN 1.3 only).
For LIN 1.3 a wakeup signal is indicated by WAKE in blue. If the wakeup signal
is not followed by a valid wakeup delimiter a wakeup error is detected and
displayed as WUP in red.
Interpreting LIN Lister Data
In addition to the standard Time column, the LIN Lister contains these columns:
ID — frame ID.
Data — (LIN 1.3 only) data bytes.
Checksum — (LIN 1.3 only).
Data and Checksum — (LIN 2.0 only).
Errors — highlighted in red.
Aliased data is highlighted in pink. When this happens, decrease the horizontal
time/div setting and run again.