Math Waveforms 4
Keysight InfiniiVision 2000 X-Series Oscilloscopes User's Guide 77
• y = channel 1, 2, 3, or 4, or g(t) (internal arithmetic operation).
• co = arbitrary constant.
• i = data point index.
The integrate operator provides an Offset softkey that lets you enter a DC offset
correction factor for the input signal. Small DC offset in the integrate function
input (or even small oscilloscope calibration errors) can cause the integrate
function output to "ramp" up or down. This DC offset correction lets you level the
integrate waveform.