148 Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide
14 Cursors
 
• Hex — Logic levels of displayed waveforms at the current X1 and X2 cursor 
positions are displayed in hexadecimal.
Manual and Track Waveform modes can be used on 
waveforms that are displayed on the analog input 
channels (including math functions).
In Hex and Binary modes, a level can be displayed as 1 
(higher than trigger level), 0 (lower than trigger level), 
indeterminate state ( ), or X (don't care).
In Binary mode, X is displayed if the channel is turned off.
In Hex mode, the channel is interpreted as 0 if turned off.
4 Press Source (or X1 Source, X2 Source in the Track Waveform 
mode); then, select the input source for cursor values.
5 Select the cursor(s) to be adjusted:
• Push the Cursors knob; then, turn the Cursors knob. To 
finalize your selection, either push the Cursors knob 
again or wait about five seconds for the popup menu to 
disappear.
Or:
• Press the Cursors softkey; then, turn the Entry knob.
The X1 X2 linked and Y1 Y2 linked selections let you adjust both cursors at the 
same time, while the delta value remains the same. This can be useful, for 
example, for checking pulse width variations in a pulse train.
The currently selected cursor(s) display brighter than the other cursors.
6 To change the cursor units, press the Units softkey.