3 Sequencing
112 Keysight M8195A Revision 2 – Arbitrary Waveform Generator User’s Guide
3.2 Sequencing Hierarchy
3.2.1 Segment
A waveform segment consists of a defined number of samples, which are played, in a
consecutive order. It is treated as a unit and can be repeated a specified number of
times or can run continuously. The sample count of segments must be in multiples of
the segment granularity. A minimum length is also required (see datasheet of the
instrument).
A segment can be played standalone (see Arbitrary Mode) or can be part of
a sequence.
Multiple segments can be combined to a sequence. A sequence can be executed
continuously or for a specified number of times.
Segment 1 Segment 2 Segment 3
Start
Trigger
Repeat Count 1 Repeat Count 2 Repeat Count 3
Sequence
Sequence Repeat Count (continuously or “n” times)
A sequence can be played standalone (see Sequence Mode) or can be part of a
scenario.