Sequencing 3
Keysight M8195A Revision 2 – Arbitrary Waveform Generator User’s Guide 119
3.6.2.3 Dynamic Select
The instrument provides a dynamic sequencing mode, which allows changing the
actually running segment or sequence without stopping and reprogramming the
instrument. The selected sequencer index is modified either by the external DYNAMIC
CONTROL input of the M8197A in multi module configuration or via remote
programming. Up to 16 M sequencer table indices can be addressed.
The run input is a software button or command, which switches the instrument from
programming mode to run mode.
3.6.3 Internal Trigger Generator
The M8195A provides a configurable internal trigger generator that allows for
generation of a periodic trigger signal that is frequency locked to the clock of the
sequencer engine. In Gated mode, the internal trigger generator provides a gate with a
width of 50% of the trigger generator period.
3.6.4 Mapping External Inputs to Logical Functions
The logical functions controlling the sequencer can be connected to multiple sources.
The following table shows all possible mappings.