EasyManua.ls Logo

Keysight M8195A - Initiate:gate[:State][?] off

Keysight M8195A
268 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
General Programming 6
Keysight M8195A Revision 2 Arbitrary Waveform Generator User’s Guide 183
6.8.6 :INITiate:GATE[:STATe][?] OFF|ON|0|1
Command
:INIT:GATE:STAT[?]
Long
:INITiate:GATE:STATe[?]
Parameters
OFF | ON | 0 | 1
Parameter Suffix
None
Description
Set or query the gate mode. This command must be used together with INIT:CONT
to set the trigger mode.
0/OFF Gate mode is off.
1/ON Gate mode is on. If continuous mode is off, the trigger mode is
“gated”.
Example
Command
:INIT:GATE:STAT ON
Query
:INIT:GATE:STAT?
Table 31: Trigger mode settings
INIT:CONT
INIT:GATE
Trigger Mode
0
0
Triggered
0
1
Gated
1
0
Continuous
1
1
Continuous