Keysight M8195A Revision 2 Arbitrary Waveform Generator User’s Guide 231
6.18.3 [:SOURce]:STABle:DATA:BLOCk? <sequence_table_index>,<length>
Command
:STAB:DATA:BLOC?
Long
:STABle:DATA:BLOCk?
Parameters
<sequence_table_index>,<length>
Parameter Suffix
None
Description
This query returns the same data as the “:STAB:DATA?” query, but in IEEE binary
block format.
<sequence_table_index> index of the sequence table entry to be accessed
<length> - number of entries to be read
Example
Query
[:SOURce]:STABle:DATA:BLOC? <sequence_id>,<length>
:STAB:DATA:BLOC? 0, 6
6.18.4 [:SOURce]:STABle:SEQuence:SELect[?]
<sequence_table_index>|MINimum|MAXimum
Command
:STAB:SEQ:SEL[?]
Long
:STABle:SEQuence:SELect[?]
Parameters
<sequence_table_index>|MINimum|MAXimum
Parameter Suffix
None
Description
Select where in the sequence table the sequence starts in STSequence mode. In
dynamic sequence selection mode select the sequence that is played before the first
sequence is dynamically selected.
Example
Command
:STAB:SEQ:SEL 0
Query
:STAB:SEQ:SEL?