6 General Programming
232 Keysight M8195A Revision 2 Arbitrary Waveform Generator User’s Guide
6.18.5 [:SOURce]:STABle:SEQuence:STATe?
Command
:STAB:SEQ:STAT?
Long
:STABle:SEQuence:STATe?
Parameters
None
Parameter Suffix
None
Description
This query returns an integer value containing the sequence execution state and the
currently executed sequence table entry
Table 43: Returned sequence state
Bit
Width
Meaning
31:27
5
Reserved
26:25
2
Sequence execution state
0: Idle
1: Waiting for Trigger
2: Running
3: Waiting for Advancement Event
24:0
25
Index of currently executed sequence table entry. In Idle state the value is
undefined.
Example
Query
:STAB:SEQ:STAT?

Table of Contents