General Programming 6
Keysight M8195A Revision 2 Arbitrary Waveform Generator User’s Guide 197
6.11.5 :INSTrument:DACMode[?] SINGle|DUAL|FOUR|MARKer|DCDuplicate|DCMarker
Command
:INST:DACM[?]
Long
:INSTrument:DACMode[?]
Parameters
SINGle|DUAL|FOUR|MARKer|DCDuplicate|DCMarker
SINGle Channel 1 can generate a signal
DUAL Channels 1 and 4 can generate a signal, channels 2 and 3 are
unused
FOUR Channels 1, 2, 3, and 4 can generate a signal
MARKer Channel 1 with two markers output on channel 3 and 4
DCDuplicate dual channel duplicate: Channels 1, 2, 3, and 4 can generate
a signal. Channel 3 generates the same signal as channel 1. Channel 4
generates the same signal as channel 2.
DCMarker dual channel with marker: Channels 1 and 2 can generate a
signal. Channel 1 has two markers output on channel 3 and 4. Channel 2
can generate signals without markers.
Parameter Suffix
None
Description
Use this command or query to set or get the operation mode of the DAC. The value
of the operation mode determines, to which channels waveforms can be transferred
and the format of the waveform data. In operation mode SINGle, DUAL, DCDuplicate,
or FOUR the data consists of 1-byte waveform samples only. In operation mode
MARKer or DCMarker the data loaded to channel 1 consists of interleaved 1-byte
waveform and 1-byte marker samples (see section :TRACe Subsystem). In operation
mode DDUPlicate waveforms can only be loaded to channels 1 and 2.
Example
Command
:INST:DACM DUAL