6 General Programming
182 Keysight M8195A Revision 2 Arbitrary Waveform Generator User’s Guide
6.8.5 :INITiate:CONTinous[:STATe][?] OFF|ON|0|1
Command
:INIT:CONT:STAT[?]
Long
:INITiate:CONTinuous:STATe[?]
Parameters
OFF | ON | 0 | 1
Parameter Suffix
None
Description
Set or query the continuous mode. This command must be used together with
INIT:GATE to set the trigger mode.
0/OFF Continuous mode is off. If gate mode is off, the trigger mode is
“triggered”, else it is “gated”.
1/ON Continuous mode is on. Trigger mode is “automatic”. The value of
gate mode is not relevant.
Example
Command
:INIT:CONT:STAT ON
Query
:INIT:CONT:STAT?