EasyManua.ls Logo

Keysight M8195A - Output[1|2|3|4]:Filter:frate[:Value][?]; Output[1|2|3|4]:Filter:frate:type[?] Lowpass|Zoh|User

Keysight M8195A
268 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
6 General Programming
210 Keysight M8195A Revision 2 Arbitrary Waveform Generator User’s Guide
Table 32: Differential offset
Value
Normal Output
Complement Output
< 0
Offset decreased
Offset increased
0
No offset
> 0
Offset increased
Offset decreased
Due to the use of DAC values, the granularity is 1.
6.13.7 :OUTPut[1|2|3|4]:FILTer:FRATe[:VALue][?]
Command
:OUTP:FILT:FRAT[?]
Long
:OUTPut:FILTer:FRATe[?]
Parameters
<value0>, <value1>…<value15> |<block>
Parameter Suffix
None
Description
Set or get the FIR filter coefficients for a channel to be used when the Sample Rate
Divider for the Extended Memory is 1. The number of coefficients is 16 and the values
are doubles between -2 and 2. They can be given as a list of comma-separated
values or as IEEE binary block data of doubles.
The coefficients can only be set using this command, when the predefined FIR filter
type is set to USER.O
Example
Query:
OUTP:FILT:FRAT?
6.13.8 :OUTPut[1|2|3|4]:FILTer:FRATe:TYPE[?] LOWPass|ZOH|USER
Command
:OUTP:FILT:FRAT:TYPE[?]
Long
:OUTPut:FILTer:FRATe:TYPE[?]
Parameters
LOWPass|ZOH|USER
Parameter Suffix
None

Table of Contents