General Programming 6
Keysight M8195A Revision 2 Arbitrary Waveform Generator User’s Guide 233
6.18.6 [:SOURce]:STABle:DYNamic:[STATe][?] OFF|ON|0|1
Command
:STAB:DYN[?]
Long
:STABle:DYNamic[?]
Parameters
OFF | ON | 0 | 1
Parameter Suffix
None
Description
Use this command to enable or disable dynamic mode.
If dynamic mode is switched off, segments or sequences can only be switched in
program mode, that is signal generation must be stopped. In arbitrary mode use
TRACe[1|2|3|4]:SELect to switch to a new segment. In sequence mode use
[:SOURce]:STABle:SEQuence:SELect to switch to a new sequence.
If dynamic mode is switched on, segments or sequences can be switched
dynamically when signal generation is active. The next segment or sequence is either
selected by the command [:SOURce]:STABle:DYNamic:SELect or by a signal fed into
the dynamic port of the M8197 module. The external input values select sequence
table entries with corresponding indices.
Example
Command
:STAB:DYN 0
Query
:STAB:DYN?
6.18.7 [:SOURce]:STABle:DYNamic:SELect <sequence_table_index>
Command
:STAB:DYN:SEL
Long
:STABle:DYNamic:SELect
Parameters
<sequence_table_index>
Parameter Suffix
None
Description
When the dynamic mode for segments or sequences is active, set the sequence table
entry to be executed next..
Example
Command
:STAB:DYN:SEL 0