6.7.10 Run Status Subsystem
The Run Status register contains the run status conditions of the individual channels.
The following SCPI commands and queries are supported:
:STATus:OPERation:RUN[:EVENt]?
:STATus:OPERation:RUN:CONDition?
:STATus:OPERation:RUN:ENABle[?]
:STATus:OPERation:RUN:NTRansition[?]
:STATus:OPERation:RUN:PTRansition[?]
Table 30: Run status register
Bit Number
Decimal Value
Definition
0 Run Status
1
Indicates if channel 1 is running
1 Run Status
2
Indicates if channel 2 is running
2 Run Status
4
Indicates if channel 3 is running
3 Run Status
8
Indicates if channel 4 is running