Sequencing 3
Keysight M8195A Revision 2 Arbitrary Waveform Generator User’s Guide 121
3.7 Sequencer Execution Flow
Programming
Waveform Download
Trigger Mode
Armed
Start First
Segment / Sequnce
Armed
Gate (Rising)
Enable
Armed
Self Armed
Y
N
Continuous
Gated
Triggered
Y
N
Self Armed
Armed
Enabled
Y
N
Armed
Trigger (Edge)
Self Armed
Armed
Enabled
Y
N
Segment / Sequence Generation
Y
N
Gate (Falling), End of Sequence (Triggered)
Stopped
RUN
Adv Event
Adv Event
Figure 30: Sequencer execution flow