6 General Programming
212 Keysight M8195A Revision 2 Arbitrary Waveform Generator User’s Guide
Example
Command:
OUTP:FILT:FRAT:DEL 10ps
6.13.11 :OUTPut[1|2|3|4]:FILTer:HRATe[:VALue] [?]
Command
:OUTP:FILT:HRAT[?]
Long
:OUTPut:FILTer:HRATe[?]
Parameters
<value0>, <value1>…<value31>|<block>
Parameter Suffix
None
Description
Set or get the FIR filter coefficients for a channel to be used when the Sample Rate
Divider for the Extended Memory is 2. The number of coefficients is 32 and the values
are doubles between -2 and 2. They can be given as a list of comma-separated
values or as IEEE binary block data of doubles.
The coefficients can only be set, when the predefined FIR filter type is set to USER.
Example
Query:
OUTP:FILT:HRAT?
6.13.12 :OUTPut[1|2|3|4]:FILTer:HRATe:TYPE[?] NYQuist|LINear|ZOH|USER
Command
:OUTP:FILT:HRAT:TYPE[?]
Long
:OUTPut:FILTer:HRATe:TYPE[?]
Parameters
NYQuist|LINear|ZOH|USER
Parameter Suffix
None
Description
Set or get the predefined FIR filter type for a channel to be used when the Sample
Rate Divider for the Extended Memory is 2.
NYQuist Nyquist filter (half-band filter) with rolloff factor 0.2
LINear Linear interpolation filter
ZOH Zero-order hold filter
USER User-defined filter
The command form modifies the FIR filter coefficients according to the set filter type,
except for type USER.