Use this tab to configure the FIR filter coefficient values for Channel 1, Channel 2,
Channel 3, and Channel 4. The number of coefficients depends on the extended
memory sample rate divider (see Sample Rate Divider ‘extended memory’ in
Output tab).
For sample rate divider 1, 2, and 4 the number of coefficients are 16, 32, and 64,
respectively.
For complete details, refer to the section “Theory of Operation”.