This tab has the following configurable fields:
Arm mode
Armed – Signal generation starts when an “enable” event is received as
defined by the trigger mode.
Self – Signal generation starts as defined by the trigger mode.
Trigger mode
Continuous – Signal generation starts immediately after pressing the Run
button. No trigger needed.
Triggered – Signal generation starts after a trigger is received.
Gated – Signal generation starts when a rising edge is received on the trigger
input and pauses when a falling edge is received. Signal generation restarts
after the next rising edge.
Threshold
Specifies the threshold voltage for a software trigger or event.
Polarity
Specifies the polarity for a software trigger or event viz. Negative, Positive, or
Either.
Operation
Specifies whether the trigger or event operation is Synchronous or
Asynchronous. Operation mode is same for both trigger and event input.
Frequency
Specifies the frequency for internal trigger.
Force Trigger
Use this button to send a software trigger to a channel.
Force Event
Use this button to send a software event to a channel.
Force Enable
Use this button to send a software “enable” to a channel.