6 General Programming
258 Keysight M8195A Revision 2 Arbitrary Waveform Generator User’s Guide
6.21.21 :TRAC[1|2|3|4]:ADV[?]
Command
:TRAC[1|2|3|4]:ADV[?]
Long
:TRACe[1|2|3|4]:ADVance[?]
Parameters
AUTO | COND | REP | SING
Parameter Suffix
None
Description
Use this command or query to set or get the advancement mode for the selected
segment. The advancement mode is used, if the segment is played in arbitrary mode.
The command has only effect, If the channel is sourced from Extended Memory. In
this case the same value is used for all other channels sourced from Extended
Memory.
Example
Command
:TRAC:ADV AUTO
Query
:TRAC:ADV?
6.21.22 :TRAC[1|2|3|4]:COUN[?]<count>|MINimum|MAXimum
Command
:TRAC[1|2|3|4]:COUN[?]
Long
:TRACe[1|2|3|4]:COUNt[?]
Parameters
<count>|MINimum|MAXimum
Parameter Suffix
None
Description
Use this command or query to set or get the segment loop count for the selected
segment. The segment loop count is used, if the segment is played in arbitrary mode.
<count> 1..4G-1: number of times the selected segment is repeated.
The command has only effect, If the channel is sourced from Extended Memory. In
this case the same value is used for all other channels sourced from Extended
Memory.
Example
Command
:TRAC:COUN 1
Query
:TRAC:COUN?

Table of Contents