7.1.2 Test inhibit
The instrument inhibits testing if it detects a circuit voltage higher than the voltage it is applying to carry out the test.
There are no disabled options for the continuity inhibit function.
The screen will display a live circuit warning if voltage is detected and make a loud, fast beeping sound.
This mode cannot be overridden and inhibits all testing until the source of the voltage is removed.
7.1.3 Hot keys for continuity checking and resistance measurement
When the MFT-X1 is being used to check continuity or measure resistance, the functions of the Hot keys are:
Hot key 1 Hot key 2 Hot key 3 Hot key 4 Hot key 5
Fig 13: Continuity Hot keys
Hot key Function Description
1 Current value
Press to select 200 mA test current, press again to return to 10 mA test current
2 Help
Press for help on continuity checking and resistance measurement
3 Save
Press to send the results to a paired Bluetooth
®
device,
Refer to 16. Download results to a remote device on page 89.
4 Buzzer On/Off
Press to turn buzzer on, press again to turn buzzer off
5 Direction
Press repeatedly to switch between unidirectional +ve to –ve and
bi-directional modes