15.5.11 Network connection
This is the ON/OFF for Bluetooth
®
communications. When switched to ON a Bluetooth
®
icon appears in the bottom
information bar and flashes ON/OFF to indicate the instrument has active Bluetooth
®
.
This must be set to ON before an instrument can be connected to a mobile device.
1. Select the network connection option.
If this option is on, the instrument can send results to a connected device via Bluetooth
®
. If it is off, the
instrument cannot communicate via Bluetooth
®
.
2. Press Hot key 3 to turn network connection off.
3. Press Hot key 3 again to turn network connection back on.
4. Your choice is stored automatically.
15.5.12 Clear network connection
The MFT-X1 does not need to be connected to a mobile device for it to connect to the same device. However, the
MFT-X1 can be connected to a mobile device if the operator is not familiar with the connection process between the
MFT and the mobile device. THis is most likely if the instrument is a pool instrument used by several people.
If there are problems connecting the MFT to a mobile device the operator can clear any connection that the
instrument may have recorded. To do this:
1. Select the Clear Network connecting option
2. Press the green check button (Hot key 3).
3. Confirm the connections are to be cleared by pressing the green check button (Hot key 4)
4. All Bluetooth
®
connections will be removed.
15.5.13 Reset to factory settings
1. Select the reset to factory settings option.
2. A warning screen appears asking if you want to proceed.
3. If you do NOT want to proceed, press Hot key 2.
4. If you DO want to reset to factory settings, press Hot key 4 and hold it for 3 seconds.
5. A progress bar appears on Hot key 4 during the 3 second countdown.
6. At the end of 3 second, all the settings and data in your instrument are erased, and all factory default settings are
restored.
www.megger.com
MFT-X1
82